Model Number ORG-9100A
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2021
Event Type  malfunction  
Manufacturer Narrative
The customer reported signal loss for 7 transmitters/rooms.According to the customer there was a signal loss error message at cns and rns for the transmitters (not simultaneously).Some transmitters were down for seconds while others took minutes for signal loss to clear.The patients were stationary so transmitters did not move to different area.The issue was resolved without user interaction.Technical service (ts) advised the customer that signal loss is when the receiver is not getting a readable signal on that frequency that the channel is programmed into.Ts advised the customer to change the channel on the org receiver to see if issue is related to the org receiver or the transmitter.There was no patient injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported signal loss for seven (7) transmitters/rooms at the central nurse's station (cns) and the remote nurse's station (rns).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9100A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
MDR Report Key11735525
MDR Text Key248261627
Report Number2080783-2021-00247
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103883
UDI-Public04931921103883
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9100A
Device Catalogue NumberORG-9100A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/27/2021
Distributor Facility Aware Date03/29/2021
Device Age73 MO
Event Location Hospital
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-