| Model Number |
750190 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cuff on a smiths medical trach tube would only partially inflate.Installation failure.They had to use another canula.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: returned device was received for evaluation.During the evaluation of the device the customer reported condition was not confirmed.No fault found.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|