TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation - practice manager.Based on the results of our investigation, the root cause of the complaint could not be identified related to our product or process.Reference photo of actual sample showed 1 piece safety needle wherein safety sheath was not activated.Photo of unit box was also provided however unit box number cannot be confirmed.Retention samples were visually confirmed free from defects that will affect activation of safety sheath and passed evaluation for sheath activation and deactivation.Also, no irregularity was encountered during the simulation of manual sheath activation that may lead to the complaint.We have series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems and needle sticks.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.(b)(4).
 
Event or Problem Description
The user facility reported the mckesson needle would not close all the way.A nurse suffered a needle stick due to the needle not closing all the way.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE
Common Device Name
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
MDR Report Key11736756
Report Number3003902955-2021-00009
Device Sequence Number14334605
Product Code MEG
UDI-Device Identifier30612479178886
UDI-Public30612479178886
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K051865
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberN/A
Device Catalogue Number102-N25105S
Device Lot Number200330B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2021
Initial Report FDA Received Date04/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-