Model Number 0998-00-0800-52
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and was able to verified the reported issue.The fse troubleshoot and found the problem to be the power management board.The problem was not corrected yet and since there is no service order report available, a supplemental report will be submitted when this information is provided to us.
 
Event Description
It was reported that during charging and checking the battery in the cardiosave intra-aortic balloon pump (iabp) will not charge.The unit charging led indicator intermittently turns on and off every 20 or 30 seconds.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, b6, b7, d10, e1(event site email), g2, g3, g6, g7, h2, h6(investigation type & investigation conclusions), h10, h11.Corrected fields: b3, d1, d4(model#), d5, g1(contact person), h4 *the aware date reported in the initial report was submitted incorrectly.The aware date should read: 03/31/2021.Good faith efforts (gfes) have been made in attempt to obtain more information.If any pertinent information is received in the future, the complaint will be reopened and updated accordingly.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11736840
MDR Text Key247689296
Report Number2249723-2021-00879
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/05/2021
Initial Date FDA Received04/28/2021
Supplement Dates Manufacturer Received01/24/2023
Supplement Dates FDA Received01/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-