| Model Number |
72200568S |
| Medical Device Problem Codes |
Overheating of Device (1437); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the light source had an e12 error.No case involved; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event, include overheating of the power supply caused by poor ventilation due to external obstructions to vents or excessive dust buildup on rear fans or fan guards.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future occurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.The complaint of e12 error could not be confirmed.Product failed to power up.Cause of power failure is a defective power supply.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|