Model Number 7211004
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during the rotator cuff repair procedure, staff found a hair mixed and enclosed in the sterilization tray when the package of the dyonics 25 inflow only tube set was opened.A backup device was available to complete the procedure with no significant delay or patient injuries.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was returned to the designated complaint unit for independent evaluation.There was no relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.The tube set was not returned with the tray.Functional evaluation of the device did not reveal any malfunction.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the lot did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 25 INFLOW ONLY TUBE SET (3)
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11737192
Report Number1643264-2021-01584
Device Sequence Number19714788
Product Code HRX
UDI-Device Identifier03596010543561
UDI-Public03596010543561
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number7211004
Device Catalogue Number7211004
Device Lot Number50908047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2021
Initial Date Received by Manufacturer 04/07/2021
Supplement Date Received by Manufacturer05/28/2021
Initial Report FDA Received Date04/28/2021
Supplement Report FDA Received Date05/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-