Catalog Number 3910-500-312
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the inserter tip broke and remained in the patient.
 
Event or Problem Description
It was reported that the inserter tip broke and remained in the patient.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: inserter tip broken probable root cause: design -inserter/implant not compatible with guide.-inserter material/design too weak.-inserter shaft cannot bend around curved guide.-drill not designed to center hole with respect to guide.-guide mouth not designed to grip bone.-guide not able to be gripped tightly.Process -inserter, implant, and/or guide manufactured out of spec.-anchor coating process out of specification.Application -excessive force impacting inserter.-movement of guide out of orientation to pilot hole.-use of wrong drill.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICONIX 1 TT WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11737212
Report Number0002936485-2021-00229
Device Sequence Number3591346
Product Code MBI
UDI-Device Identifier37613327176927
UDI-Public37613327176927
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K170098
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-500-312
Device Lot Number20347AE2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/31/2021
Supplement Date Received by Manufacturer03/31/2021
Initial Report FDA Received Date04/28/2021
Supplement Report FDA Received Date06/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age17 YR
-
-