MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER
|
Back to Search Results |
|
Model Number PED2-425-35 |
Device Problems
Activation Failure (3270); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received a report that the marksman microcatheter was introduced and placed distally in the middle cerebral artery (mca).The pipeline was prepared and purged, and then the stent was delivered to the distal part of the marksman microcatheter.However, it was very difficult to keep pushing as the stent stuck to the microcatheter and could not be pushed and released.It was also reported the pipeline had failed to open in the distal segment.Less than 50% of the stent had been deployed, re-sheathing was performed less than three times, and no additional steps were taken in an attempt to open the device.The stent could not be removed, so it was necessary to completely extract the marksman microcatheter with the pipeline.Another marksman and another stent were used to complete the procedure.The patient was undergoing surgery for treatment of an amorphous aneurysm of the right ophthalmic segment.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet therapy (dapt) was not administered.
|
|
Manufacturer Narrative
|
E1: street 1 (cont.): (b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received reported that the pipeline was not located in a bend when it failed to open.There was no damage to the pipeline or catheter observed.
|
|
Manufacturer Narrative
|
G4: pma # p100018.S026 was omitted from previous report in error.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: based on the device analysis and reported information, the customer¿s report of ¿failure/incomplete open distal¿ was unable to be confirmed as both braid ends were found to be opened.It is likely the failure to open is the result of vessel tortuosity.Furthermore, the review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Based on the device analysis and reported information, the customer¿s report of ¿resistance/stuck during delivery¿ was unable to be confirmed as the pipeline flex returned without the marksman catheter used during the event.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary (see tr-nv12798 - pipeline-catheter interaction and capa2015-010156).Based on the investigation conducted resistance can occur during tracking, deployment and re-sheathing of the device in distal and tortuous anatomies.In addition, resistance can occur due to patient vessel tortuosity, failure to maintain a continuous flush, or pipeline is pulled back/torqued during delivery.The root cause could not be determined.Based on the device analysis and reported information, the customer¿s report of ¿catheter resistance¿ was confirmed as the pipeline was unable to be removed from the marksman catheter.It is likely the damages to the catheter body (accordioned) contributed to the resistance.Possible contributing factors to ¿catheter resistance¿ include the use of an incompatible device, severely tortuous patient anatomy, and insufficient catheter flush.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|