BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7010
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was a complete separation at 86.6cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 13apr2021.It was reported that shaft kink occured.The 93% stenosed, 14mm in length target lesion was located in the moderately tortuous and moderately calcified, 2.5mm in diameter left anterior descending artery.A 2.5mm x 12mm quantum maverick balloon catheter was advanced for dilation.During the procedure, it was noted that the shaft was kinked 85cm away from the distal end of the hub.The procedure was completed with another of same device.There were no patient complications reported and the patient was stable.However, returned device analysis revealed a complete separation at 86.6cm distal of the strain relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11738721
MDR Text Key247931277
Report Number2134265-2021-05412
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392651
UDI-Public08714729392651
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/11/2022
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0024917947
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/13/2021
Initial Date FDA Received04/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient Weight63
-
-