DATASCOPE CORP. - FAIRFIELD TRANS-RAY PLUS 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0605
Device Problems Filling Problem (1233); Calibration Problem (2890)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  malfunction  
Manufacturer Narrative
Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console indicated that catheter inspection was required.It was also indicated that there was a leak in the iab circuit, blood detected, an auto-fill failure, and the optical sensor was unable to calibrate.The iab was removed.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
 
Event Description
N/a.
 
Manufacturer Narrative
The product was returned with the membrane completely unfolded with blood traces only on the statguard.The returned maquet sheath was located at 35.3cm from the iab tip.The stopcock, extender tubing and pressure tubing were also returned.An inner lumen kink, catheter tubing kink and optical fiber breakage were observed at 35.3cm from the iab tip.A leak test of the balloon, catheter, pressure tubing, y-fitting, extracorporeal tubing and extender tubing was performed and no leaks were detected.A guidewire was then inserted into the inner lumen.There was a smooth insertion with an obstruction at the inner lumen kink location and blood traces were found on the guidewire.The iab was placed on a cs300 pump for two hours which represents one complete autofill cycle.The iab pumped normally, no alarm sounded.The optical fiber was found to be broken, confirming the reported failure "difficult/unable to calibrate optical sensor".However, the reported alarms "autofill failure, "check iab catheter", "blood detected" and "leak in circuit" were not confirmed as there were no leaks and the iab pumped normally.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period may-2019 through apr-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key11738731
MDR Text Key247898869
Report Number2248146-2021-00269
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2023
Device Catalogue Number0684-00-0605
Device Lot Number3000127406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2021
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CS300 SI251176H4).
-
-