Model Number FF562R |
Device Problems
Break (1069); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturer evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
|
Event Description
|
It was reported to aesculap inc.That a caspar rongeurstrserr 2x12mm185mm (part # ff562r) was used during an unknown procedure performed on (b)(6) 2021.According to the complainant, during the procedure, a piece of the instrument broke away and detached into the patient.The fragment was retrieved under flouroscopy.The complaint device was available to be returned to the manufacturer for evaluation.No patient complications were reported as a result of the fragment retrieval.Although requested, additional information has not been made available.The adverse event / malfunction is filed under aag reference xc (b)(4).
|
|
Event Description
|
Update: the following additional information was received via mw5100488: the event date was noted as (b)(6) 2021 (previously reported as (b)(6) 2021).The malfunction occurred during an l3-s1 laminectomy procedure.No harm to patient and no surgical delay.
|
|
Event Description
|
No updates required.
|
|
Manufacturer Narrative
|
Manufacturer site evaluation: the complaint device was returned to the manufacturer for evaluation.A visual examination was performed which found that the jaw had broken off.A deformation to the jaw hinge was identified.Additionally, an optical inspection of the fracture surface for the broken off jaw was performed.No anomalies or deviations were identified.The device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.The investigation into the cause of the reported problem was able to confirm the failure mode.The device jaw had broken away.Although a root cause was not able to be identified, the investigation determined that there was no indication for a material-, manufacturing- or design-related failure.
|
|