Model Number 40456
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the tip of the wire broke off and was snared.The target lesion was located in the mildly calcified left superficial femoral artery.A 300cm v-14 control wire was used together with a rubicon guide catheter.During withdrawal, after balloon angioplasty, there was trouble removing the device as the distal tip of the wire got stuck.A rubicon was used to unfold the device but the tip broke off.Subsequently, the rubicon and some of the wire was removed.The fractured part of the wire was snared.Per physician's opinion, the wire was defective and that it was not being forced in anyway.The procedure was completed successfully.No further complications reported and the patient has fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-14 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11739690
Report Number2134265-2021-05314
Device Sequence Number17204632
Product Code DQX
UDI-Device Identifier08714729799238
UDI-Public08714729799238
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K112745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2023
Device Model Number40456
Device Catalogue Number40456
Device Lot Number0026631470
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/07/2021
Initial Report FDA Received Date04/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-