| Model Number |
MF055R |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: additional information / investigation results will be provided in a supplemental report, if available.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap inc.That a decker rongeurs 2x6mm str 152mm (part # mf055r) was used during lumbar discectomy and foraminotomy procedure performed on (b)(6) 2021.According to the complaint description, the surgeon was using the instrument during the case and the tip broke off and was retained in the patient.There was a 30 min delay in the surgery.Further information was received on apr 13, 2021.Additional intervention was required, longer procedure time and x-ray time for the surgeon to try and recover the broken piece.The broken piece is still in the patient.The patient outcome is unknown.The complaint device was returned to manufacture for evaluation.The adverse event is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the device was sent back to the manufacturer for further evaluation.They conducted dimensional, visual, hardness, and functional testing on the device.All tests determined, that the device was conforming to its specifications.A definitive root cause was not able to be determined.Potential causes for the tip breaking off could be excessive force or torquing applied to the working end of the device by the user.Corrective action: corrective or preventive action was not deemed necessary, since the error was not attributed by a manufacturing error.And there have been no previous occurrences.
|
| |
|
Search Alerts/Recalls
|