MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 199723860S
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Codes Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional product code : kwp, kwq.Complainant part is expected to be returned for manufacturer review/ investigation but has yet to be received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported on an unknown date that the patient was found that the screws which were deployed in the right ilium had come off the rod.He underwent for a revision surgery for rod fusion, and it was confirmed that remaining screws had not been loose.The plf revision procedure was performed to extend the fusion range up the ilium on (b)(6) 2021.The procedure was completed less than 30-minute delay.The patient outcome was unknown.Concomitant device reported: unknown set screw (part # unknown, lot # unknown, quantity unknown), unknown tightener (part # unknown, lot # unknown, quantity unknown), unknown rod holder (part # unknown, lot# unknown, quantity unknown), 5.5 exp verse unitized set scr (part # 199721001s, lot # xp1233, quantity 1), single-inner set screw (part # 179702000, lot # 286895, quantity 1).This complaint involves four (4) devices.This report is for (1) 5.5 exp verse fen scr 8.0 x 60.This report is 3 of 3 (b)(4).Related product complaint: (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h11: d1; brand name.D2: additional product code : osh, mni, mnh.D4: part number, lot# & udi updated.G1: manufacture site.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D1; brand name.D2: additional product code : osh, mni, mnh.D4: part number, lot# & udi updated.G1: manufacture site.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to us cq for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that single-inner setscrew component threads were slightly deformed.The dimensional inspection was not performed due to the post-manufacturing damage.The complete functional test cannot be performed as all mating devices were not returned and components in the received device were damaged.The observed condition of the single-inner setscrew component in the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed deform condition of the single-inner setscrew component would contribute to the complained loose issue.While no definitive root cause could be determined, it is probable that the single-inner setscrew is deformed due to the unintended.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: the dhr of product code: 179702000 lot : 286895 was electronically reviewed and no non-conformances were observed during the manufacturing process.The product was released on: 29.09.2020 qty: 1335 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G4.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11740288
Report Number1526439-2021-00818
Device Sequence Number14397362
Product Code NKB
UDI-Device Identifier10705034444291
UDI-Public10705034444291
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number199723860S
Device Catalogue Number179702000
Device Lot Number286895
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2021
Initial Date Received by Manufacturer 04/02/2021
Supplement Date Received by Manufacturer04/30/2021
04/30/2021
06/07/2021
Initial Report FDA Received Date04/28/2021
Supplement Report FDA Received Date05/27/2021
05/28/2021
06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE FEN SCR 8.0X60.; SINGLE-INNER SETSCREW.; UNKNOWN LOCKING/SET SCREWS.; UNKNOWN ROD HOLDER.; UNKNOWN RODS.; UNKNOWN TIGHTENER.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-