BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 24 IN. TOUCH SCREEN; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10003
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during assembly of cori, the 24 inch touchscreen monitor will turn on, yet nothing will appear on the screen.When they attempted to press "menu", nothing happened, and the monitor turns off on it¿s own.They ensured everything was plugged in appropriately (the tablet works and portrays what should be on the monitor screen), but they don't see anything on the monitor.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the cori touchscreen, p/n rob10003, (b)(6), intended for used in treatment, was returned for evaluation.A relationship between the reported event and the device was not establish.The touchscreen pass all testing, no malfunction was observe.Touchscreen operated as design.In addition, a functional evaluation was performed on the part beyond the reported complaint.No additional functional non-conformances were identified.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, factor that may have contributed to the reported symptom may have been associated with internment electrical failure.A historical capa, nc, hhe/pra, field action review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and that the documented risk level is still adequate.No further containment or corrective action is recommended or required at this time, all complaints are monitored through complaint investigation and post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE 24 IN. TOUCH SCREEN
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11741223
Report Number3010266064-2021-00319
Device Sequence Number16933412
Product Code OLO
UDI-Device Identifier00885556756256
UDI-Public00885556756256
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10003
Device Catalogue NumberROB10003
Device Lot NumberGMO312F040003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2021
Supplement Date Received by Manufacturer02/01/2022
Initial Report FDA Received Date04/28/2021
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-