• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ARTHROSCOPE Back to Search Results
Model Number 375-545-000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the inner shaft tip was completely detached from the shaft and not returned with the complaint device.A review of the dhr could not be performed as the lot number was not reported.The most likely root causes are the user applied too much force to the device or the arthroscopic shaver may have come into contact with staples, clips, or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturers¿ manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Follow a suitable surgery protocol.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the tip of the cutting blade of the arthroscopic shaver broke off inside the patient.The broken tip was retrieved arthroscopically using a grasper without x-ray or a larger incision.The complainant is not aware of any adverse consequences.No harm to the patient was noted.There was no unanticipated medical intervention and the extended procedure time was approximately 15 minutes.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key11742271
Report Number0001056128-2021-00022
Device Sequence Number19605832
Product Code HRX
UDI-Device Identifier00885825014223
UDI-Public00885825014223
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K012635
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number375-545-000
Device Catalogue Number375-545-000RR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2021
Initial Report FDA Received Date04/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-