• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN RK HK OFFSET COUPLER TRIAL STD 4MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN RK HK OFFSET COUPLER TRIAL STD 4MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71431909
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection for a total knee arthroplasty the legion revision hinge offset coupler trial standard 4 mm was noticed with a bio burden.Smith and nephew backup device available.No delay or injury reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN RK HK OFFSET COUPLER TRIAL STD 4MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11742451
Report Number1020279-2021-03615
Device Sequence Number17280105
Product Code HRY
UDI-Device Identifier00885556537329
UDI-Public00885556537329
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71431909
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/08/2021
Supplement Date Received by Manufacturer05/25/2021
Initial Report FDA Received Date04/28/2021
Supplement Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-