• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L GENERAL PURPOSE TEMPERATURE PROBE; CLINICAL ELECTRONIC THERMOMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L GENERAL PURPOSE TEMPERATURE PROBE; CLINICAL ELECTRONIC THERMOMETER Back to Search Results
Model Number 81-020409
Medical Device Problem Codes Detachment of Device or Device Component (2907); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Gastritis (1874); Unspecified Gastrointestinal Problem (4491)
Date of Event 04/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation finding: manufacturing plant inspected a total of (b)(4) units and all parts were found in compliance with the process specifications.Past complaint logs have been reviewed and confirmed of no similar issues have been reported.Root cause: no root cause determined.The sus involuntary report (mw5100669) stated the following, "patient was agitated in icu and may have bitten through the wire." lot information provided by the customer does not match any of the company's device history records (does not exist).The investigation is complete at this time.No further information is available at this time.We will provide follow up report if additional information becomes available.
 
Event or Problem Description
According to the customer a piece of the temperature probe was broken off and found inside the patient stomach.Patient potentially bit off the probe per sus voluntary event report (mw5100669).The sus voluntary event report stated the following: patient complains of abdomen distention and nausea.Abdomen without peritonitis or guarding.Obstruction series shows dilated small bowl, air in colon concern for ileus.But also shows a retained object in luq that is concerning for portion of dobhoff tube (feeding tube) vs esophageal temp probe that appears to be in the stomach.Chest x-ray showed intact dobhoff coursing through esophagus into stomach gi consulted for esophagogastroduodenoscopy (egd).Egd appeared to be a fractured dobhoff tube in the body of the stomach that extended into the antrum and into the duodenum.Some gastritis with hematin likely from the trauma from the presence of a foreign body in this regin.Pathology report of a 0.3 cm in diameter rubbery tan white rubber coated wire.Sounds like the esophageal probe.Patient was agitated in icu and may have bitten through the wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENERAL PURPOSE TEMPERATURE PROBE
Common Device Name
CLINICAL ELECTRONIC THERMOMETER
Manufacturer (Section D)
DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L
global park, p.o. box 180-3006
602 parkway, la aurora
heredia 146
CS  146
MDR Report Key11744372
Report Number9613793-2021-00011
Device Sequence Number4065934
Product Code FLL
UDI-Device Identifier00749756046145
UDI-Public00749756046145
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K925791
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number81-020409
Device Lot Number223800
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2021
Initial Report FDA Received Date04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
MW5100669
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-