| Model Number |
18266 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the kpa value falls outside the limit value.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed., corrected data: d4: serial # corrected from "(b)(6)" to "(b)(6)".H4: device manufacture date corrected from "03-apr-2020" to "08-jun-2020".
|
| |
|
Event or Problem Description
|
|
The customer reported the kpa value falls outside the limit value.No patient impact or consequences were reported.
|
| |
|
Search Alerts/Recalls
|