KARL STORZ ENDOVISION CYSTO-URETHRO-FIBERSCOPE; CYSTO-URETHROSCOPE Back to Search Results
Model Number 11272C1
Medical Device Problem Codes Leak/Splash (1354); Material Integrity Problem (2978); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Urinary Tract Infection (2120); Insufficient Information (4580)
Date of Event 02/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's preliminary comments: the initial analysis of the causes of this incident: device was not yet returned for investigation; therefore, final root cause could not be identified.The pre-investigation showed that the endoscope (b)(4) was delivered to the customer on 28.08.2019.Since this time, the endoscope has been used 455 times according to the customer.Since delivery, the endoscope has never been sent in for repair or inspection.The list of similar incident that have occurred in (b)(6) or in europe with this cystoscope reference over the last 3 years: there is no similar incident that has occurred in (b)(6) or in europe over the last 3 years.The specific recommendation mentioned in the manual of this device in order to avoid the orrcurrence of the type of incident: according to ifu, the following recommendations were stated: - inspect the fiberscope prior to each surgical procedure to ensure that it is functioning correctly and that it has been correctly cleaned, disinfected and (where applicable) sterilized.- the fiberscope is delivered unsterilized, and must therefore be cleaned, disinfected and (where applicable) sterilized prior to initial use and each subsequent reuse.- several chapters in the ifu contain instructions for cleaning, disinfection, sterilization and reprocessing.See chapters "14.Cleaning, disinfection, sterilization", "15.Preliminary cleaning", "16.Manual reprocessing", "17.Machine reprocessing", "18.Inspection after reprocessing", "19.Gas sterilization/ plasma sterilization", "20.Chemical sterilization".The number of cystoscopes installed in france with this model: total of (b)(4) cysto-urethro-fiberskops installed in (b)(6).
 
Event or Problem Description
As per a manufacturer incident report we received from the factory in (b)(4): from (b)(6) 2021, hospitalization of 3 patients for klebsialla pneumoniae ebsle urinary tract infection.This report is for hospitalization of patient #1 of 3.
 
Additional Manufacturer Narrative
As per manufacturer's investigation results: the device in question was returned to the manufacturing site in germany on 02.06.2021 for investigation.Upon the evaluation it could be confirmed that the endoscope shows signs of wear, which have also led to leaks as stated in the related investigation report: the chemical attack and leakage under the kink protector is unlikely to be a possible cause of the infections.Since the kink protector has no patient contact, there is no critical consideration for this.No microorganisms migrate from the kink protection into the canal system.The root cause for corrosion is most likely, that the cleaning agents which were used for reprocessing, were too aggressive.The damage of the product is not caused by a production problem or material defect.In addition, we would like to call your attention to the description of reprocessing, as written in the ifu, chapter 14.However, according to ifu, the following recommendations were stated: inspect the fiberscope prior to each surgical procedure to ensure that it is functioning correctly and that it has been correctly cleaned, disinfected and (where applicable) sterilized.The fiberscope is delivered unsterilized, and must therefore be cleaned, disinfected and (where applicable) sterilized prior to initial use and each subsequent reuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-URETHRO-FIBERSCOPE
Common Device Name
CYSTO-URETHROSCOPE
Manufacturer (Section D)
KARL STORZ ENDOVISION
91 carpenter hill road
charlton MA 01507
MDR Report Key11744779
Report Number1221826-2021-00039
Device Sequence Number3856883
Product Code FBO
UDI-Device Identifier04048551226100
UDI-Public4048551226100
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K945185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number11272C1
Device Catalogue Number11272C1
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/19/2021
Supplement Date Received by Manufacturer03/19/2021
Initial Report FDA Received Date04/29/2021
Supplement Report FDA Received Date08/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-