| Model Number |
80617 |
| Medical Device Problem Codes |
Collapse (1099); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The device was inspected for any damage or irregularities.The renegade showed damage in the form of shaft damage from approx.2 mm to 3.6 cm from the tip.An obstruction was determined to be approx.2 cm from the tip and was removed with a guide wire.Ftir testing showed the obstruction to be similar in composition to the fluorocarbon fiber used in the inner tube.Further analysis revealed damage in the form of kinks 131.5, 141 cm and 142 cm from the hub.Inspection of the remainder of the device revealed no damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Based on device analysis completed on 05apr2021.It was reported that distal part of the catheter was bulging.A 150/30 renegade stc 18 was selected for use.Upon preparation, it was noted that the distal part of the microcatheter was obstructed.Subsequently, there was a bulging in the distal part approximately 3-4cm.No patent complications were reported.However, device analysis revealed an obstruction of foreign material.
|
| |
|