| Model Number |
NX012Z |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If additional information or investigation results become available, they will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with a as vega ps femoral component.According to the complaint description, a vega femur was revised and it was noted "no cement was found on backside of implant".Loosening was also noted.The initial surgery was (b)(6) 2017.A revision surgery was necessary.Additional information was not provided nor available / was not available.The adverse event is filed under xc reference (b)(4).Involved components: nx056z - as vega ps tibial plateau cemented t3+ - 52001076.Nx063z - as tibia extension stem 14x52mm cemented - 52066679.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Explanation and rationale: in the light of the small amount of information received and due to the circumstance that we did not receive the complained devices for investigation, it is not possible to determine a definitive root cause for the mentioned failure.It could be possible that there were problems with the cement technique and due to this the implant has loosened prematurely.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.A capa is not necessary.
|
| |
|
Event or Problem Description
|
|
Update: the loosening was noted via x-ray results dated (b)(6) 2021; the revision occurred on (b)(6) 2021.
|
| |
|