| Model Number |
M00545160 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, when the physician cut the papilla, the cutting wire of autotome rx 49 was burnt and had detached.No part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was broken and blackened.The broken end of the cutting wire and the distal end of the catheter were observed under magnification.The distal section of the broken cutting wire was bent.The dimensional and functional evaluation were not performed due to the condition of the cutting wire.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, and blackened.The distal section of the cutting wire was also bent.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been caused by the interaction with another device and/or the manner the device was handled.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, when the physician cut the papilla, the cutting wire of autotome rx 49 was burnt and had detached.No part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|