• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number BELLAVISTA 1000
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire identification number: (b)(4).At this time, the suspected device and log file has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the bellavista 1000 is making an abnormal noise during functioning, heats up a lot, and stops working.It happened during patient use with no allegation of harm.However, intervention is unknown/not specified.
 
Manufacturer Narrative
Results of investigation: analysis on the log file shows that the noise was most likely due to the obstruction on the cooling fan which was identified and resolve by the customer.According to customer issue was not reproducible.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
MDR Report Key11746602
MDR Text Key250580904
Report Number3004553423-2021-00979
Device Sequence Number1
Product Code CBK
UDI-Device Identifier07640149380019
UDI-Public(01)07640149380019
Combination Product (y/n)N
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 04/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000
Device Catalogue Number301.100.000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Initial Date Manufacturer Received 04/02/2021
Initial Date FDA Received04/29/2021
Supplement Dates Manufacturer Received06/21/2021
Supplement Dates FDA Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-