| Catalog Number |
7210569 |
| Medical Device Problem Codes |
Decrease in Pressure (1490); Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the positioner spider limb was not holding position and it was loosing pressure.No case reported; therefore, no patient was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product.No issues were noted.A functional evaluation was performed.The device failed the weight test.The piston migration was at 2.3 inches.The reported complaint of "not holding" was confirmed and the root cause has been associated with a seal failure.The reported complaint of "losing pressure" was confirmed and the root cause has been associated with a seal failure.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|