Event occurred on an unknown date in 2021.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Investigation summary: customer quality investigation: the implant(s) was not returned, and the investigation will be completed based on the supplied image(s).The image(s) was reviewed, and the complaint condition was confirmed, as the image showed a broken spacer.As the implant(s) was not returned and as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.Conclusion: there is no indication that a design or manufacturing issue contributed to the complaint as the circumstances during the time of the event are unknown.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history review: a manufacturing record evaluation could not be performed as the lot number could not be determined from the image.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|