• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. EXPRT REVISION HIP SYSTEM; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 85MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. EXPRT REVISION HIP SYSTEM; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 85MM Back to Search Results
Model Number 495-01-085
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to patient dislocated and was being worked up for possible infection.
 
Additional Manufacturer Narrative
Corrected data : see d.8.Manufacturer narrative : the reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 37 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as a dislocation and infection.The previous surgery and the surgery detailed in this event occurred 37 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to a dislocation and infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT REVISION HIP SYSTEM
Common Device Name
EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 85MM
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key11748135
Report Number1644408-2021-00389
Device Sequence Number4357530
Product Code KWZ
UDI-Device Identifier00190446153863
UDI-Public(01)00190446153863
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K161610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/19/2024
Device Model Number495-01-085
Device Catalogue Number495-01-085
Device Lot Number549U1021
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2021
Supplement Date Received by Manufacturer05/13/2021
06/08/2021
Initial Report FDA Received Date04/29/2021
Supplement Report FDA Received Date06/03/2021
06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-105, LOT 904B1919; 430-98-064, LOT 595G1057; 495-15-170, LOT 575U1016; 497-36-000, LOT 638C1256; 931-36-764, LOT 890N1919; 411-00-105, LOT 904B1919; 430-98-064, LOT 595G1057; 495-15-170, LOT 575U1016; 497-36-000, LOT 638C1256; 931-36-764, LOT 890N1919
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-