A physician reported that on (b)(6) 2021 the certas valve was implanted to a patient via v-p shunt with a setting of 1.After that, the setting was changed to 8 and underdrainage was observed.According to doctors, certas usually does not change the set pressures for mris below 3.0 tesla.However, the settings may have changed after the mri was taken on (b)(6).The valve was removed on (b)(6) 2021 and a revision surgery was performed.No further information was provided.
|
The valve was returned for evaluation.Failure analysis - the valve was visually inspected; a needle hole was noted in the needle chamber.The position of the cam when valve was received was at setting 1.The valve was hydrated.The returned catheter was flushed and no occlusion was noted.The catheter was leak-tested and no leakage was noted.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer "underdrainage, blockage" could be due to biological debris and protein build up that may cause an occlusion.But, at the time of investigation, no programming or occlusion issues were noted.
|