| Model Number |
68LP-400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/31/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported event of deflation was received on july 06, 2021 with lot number: 3216583.Visual analysis of the returned device identified: flat creases and no openings were found in the device.A leak test was performed which identified no leakage.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.Additional, changed, and/or corrected data.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|