• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 37714
Device Problems Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Charging Problem (2892)
Patient Problems Ambulation Difficulties (2544); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative (rep) regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that the ins was currently dead and overdischarged.The caller had notes that ins had been overdischarged before and provided date of 2017 specifically.The caller was going to be meeting with patient and hcp today so they could obtain further information.Additional information was received from a rep.The rep reported that the patient has no desire to go through a prm process again.It was reported that to the best of her memory, the stimulator was last on and working properly one year ago.The paresthesia in her legs made it more difficult for her to walk and that it wasn't helping as much as it did before with the pain.The recharge burden was more than she could keep up with.The healthcare provider (hcp) discussed several options for moving forward all of which would involve a new battery and the use of high frequency stimulation programmed below the level of perception.The rep noted that they would need to test impedances thoroughly due to the existing surgical lead being nearly 20 years old and cautioned against implanting a new battery is the condition of the lead was such as to not allow successful stimulation.The patient decided to move forward with replacement.No procedure had been scheduled yet. .
 
Event Description
Additional information from the manufacturer representative reported that the revision was going to be on 20-may.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE SENSOR
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key11749148
MDR Text Key262408797
Report Number3004209178-2021-06941
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00613994610430
UDI-Public00613994610430
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2013
Device Model Number37714
Device Catalogue Number37714
Was Device Available for Evaluation? No
Date Manufacturer Received05/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-