• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 81 CM LENGTH, 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 81 CM LENGTH, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number 407452
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316); Material Separation (1562)
Health Effect - Clinical Codes Stroke/CVA (1770); Foreign Body Embolism (4439)
Date of Event 03/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
One 8.5f swartz braided introducer sheath was received for evaluation.The reported event of sheath fracture and detachment was confirmed.The soft gray tip and sheath tubing had been fractured into pieces and the tip marker exposed.In addition, inspection of the device identified degradation of shaft material.The damage is consistent with the product being stored outside of the shelf box and exposed to light.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The product instructions for use (ifu) artmt100093395 ver.A warns that product should be stored in a cool, dark, dry place.The cause of the degradation is consistent with not following the instructions for use.The cause of the patient stroke and patient insertion difficulties remains unknown.
 
Event or Problem Description
During a ventricular tachycardia procedure, the gray soft tip of the sheath detached in patient and two days later, the patient experienced stroke symptoms.An mri was conducted to initially diagnose the stroke.During the procedure, difficulty was noted when inserting a non-abbott 8.5f catheter into the introducer after mapping and ablation was conducted.Additionally, the 8.5f non-abbott catheter was difficult to fully insert through the sheath and manipulating the catheter in the patient felt abnormal.The physician removed the sheath and catheter together from the patient and upon removal, the gray soft tip of the sheath was noted to have been detached.The remainder of the sheath tip was sheared and hanging at the end of the sheath.No intervention was administered as the gray soft tip pieces had appeared to disintegrate in the patient and all pieces were not accounted for since they could not be visualized due to being opaque.Two days following the procedure the patient experienced stoke symptoms of blurred vision of the left eye, however no intervention was noted to have been conducted.Further information regarding pre anticoagulation and transesophageal echocardiography and post procedure anticoagulation are unknown.It was unknown if the stroke symptoms were related to the shearing of the device.The patient's symptoms have been noted to have been mostly resolved.Prior to the procedure, the introducer was noted to have been stored on a rack outside of its original packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0, 81 CM LENGTH, 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key11749586
Report Number3005334138-2021-00235
Device Sequence Number19900576
Product Code DYB
UDI-Device Identifier05414734205764
UDI-Public05414734205764
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number407452
Device Catalogue Number407452
Device Lot Number6970231
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2021
Initial Report FDA Received Date04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-