AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40E
Medical Device Problem Codes Difficult to Insert (1316); Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient was diagnosed with aphakia in the left eye and mccannal suture pars plana approach procedure was booked.The suspect intraocular lens (iol) was inserted into patient's eye, however, did not stay in place due to lack of support as the haptics were too short.The lens was removed with no injury and replaced with another lens of different model.Additionally, a 10-0 nylon suture was used.No further information was available.
 
Additional Manufacturer Narrative
Device evaluation: product testing could not be performed since the product was not returned for evaluation.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes; section d9: returned to manufacturer on: june 8, 2021; section h3: device evaluated by manufacturer? yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed, viscoelastic residue on the optic body and haptics.Which is consistent with a lens that was handled, during explant.The lens was cleaned, a scratch was observed, on the optic body.One haptic was detached, and the other haptic was damaged (bent).Due to the condition of the returned product, no further testing can be performed.Therefore, the complaint issue could not be confirmed.And no product deficiency could be identified.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11750154
Report Number2648035-2021-07700
Device Sequence Number14532778
Product Code HQL
UDI-Device Identifier05050474502062
UDI-Public(01)05050474502062(17)210810
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2021
Device Model NumberAR40E
Device Catalogue NumberAR40E00180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/01/2021
Supplement Date Received by Manufacturer05/10/2021
06/30/2021
Initial Report FDA Received Date04/29/2021
Supplement Report FDA Received Date06/07/2021
07/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
-
-