• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SOLARA CRT-P MRI SURESCAN PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SOLARA CRT-P MRI SURESCAN PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number W1TR03
Device Problem Failure to Capture (1081)
Patient Problem Dizziness (2194)
Event Date 04/02/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented with lightheadedness. It was also reported that the cardiac resynchronization therapy pacemaker (crt-p) was programmed with subthreshold pacing outputs. When the device feature that monitors pacing thresholds at specified intervals was performed, there was loss of capture both with the support cycles and test paces. The device was reprogrammed and remains in use. No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSOLARA CRT-P MRI SURESCAN
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11750180
MDR Text Key262189563
Report Number3004209178-2021-06949
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735675
UDI-Public00643169735675
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/28/2019
Device Model NumberW1TR03
Device Catalogue NumberW1TR03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/02/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured08/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/29/2021 Patient Sequence Number: 1
-
-