• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. UNFOLDER SILVER; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. UNFOLDER SILVER; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number PSCST30
Medical Device Problem Codes Break (1069); Crack (1135); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age/date of birth, weight, ethnicity: unknown/not provided.If implanted, give date: not applicable, as the cartridge is not an implantable device.If explanted, give date: not applicable, as the cartridge is not an implantable device.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.Should additional information be received regarding this event the case will be reopened and processed accordingly.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported when inserting the intraocular lens (iol), they tried pushing the iol, and the pscst30 cartridge split at the tip and ripped the lens and haptics.The lens was not stuck in the cartridge and there was no patient injury.Another johnson and johnson lens with the same model and diopter size was implanted in the patient's ocular sinister (left eye).Patient post-op status was reported as doing fine.No further information is available.
 
Additional Manufacturer Narrative
Additional information: device evaluation: product testing could not be performed because the product was not returned.A product quality deficiency could not be confirmed.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed that four (04) additional complaint folders were found as part of this production order (po) number.Complaint 1 - no product investigation was created.Complaint 2 & 3 - product investigation is pending.Complaint 4 ¿ no product deficiency was identified conclusion: as a result of the investigation, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: upon further review, it was noted that in the initial mdr section h6, device codes device partially delivered, haptic damage, and iol torn were inadvertently not captured.Therefore, this supplemental filing is to correct the medical device problem codes there were was not originally captured.The following sections have been updated accordingly: section h6: medical device problem code: 1069 - haptic damage, 22920 - device partially delivered, 4008 - iol torn.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNFOLDER SILVER
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11750434
Report Number2648035-2021-07638
Device Sequence Number4318925
Product Code KYB
UDI-Device Identifier05050474530034
UDI-Public(01)05050474530034(17)210923(10)CH14579
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K961242
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/23/2021
Device Model NumberPSCST30
Device Catalogue NumberPSCST30
Device Lot NumberCH14579
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/01/2021
Supplement Date Received by Manufacturer05/06/2021
08/09/2021
Initial Report FDA Received Date04/29/2021
Supplement Report FDA Received Date05/26/2021
08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Z9002 SERIAL NUMBER: (B)(6).
Patient Age73 YR
-
-