The subject device was returned to olympus (b)(4).Olympus (b)(4) reported, "all contents of the initiation were reviewed with the picture of the damaged part, therefore, the actual device would not be sent to the manufacturer." the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
|
This is a supplemental report to provide additional information.The subject device was returned to olympus¿s local distributor, but not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.According to the provided photo, it was confirmed that the probe was removed.The manufacturing record was reviewed and found no irregularities.It is possible to infer the cause from the results of past similar investigations, therefore it was determined that an investigation using equipment with similar structure or similar device is unnecessary.Based on the past similar cases, omsc presumes that the event occurred due to the following occurrence mechanism.Contact with a surgical instrument: 1.During output in seal & cut mode, the probe came in contact with hard tissue, metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.3.The probe broke when added load.Contact with non-insulated area of grasping section: 1.The distal end of the tissue pad was worn away because "seal & cut" output was activated while grasping nothing (including the case the user kept activating after cutting tissue).2.The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.3."seal & cut" output was activated under this situation, then the scratches indicating that the probe and grasping section were in contact with each other were made.4.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.5.The probe broke when added load.The above device handling has warned in the instruction manual.
|