Model Number PSCST30
Medical Device Problem Codes Break (1069); Crack (1135); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Weight and ethnicity: unknown/ not provided.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that when they were trying to insert the intraocular lens (iol) into the patient's right eye, the unloader (cartridge) split at the tip and ripped the lens and haptics.It was stated that the patient was doing fine.No other information was provided.
 
Additional Manufacturer Narrative
Device evaluation: product evaluation could not be performed since at the time of this investigation, the product had not been received.Per information provided material is not available.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that four additional complaints were received from this production order.In one complaint, no investigation was requested.In two complaints, still waiting for investigation results.In last one, no product deficiency was identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNFOLDER SILVER
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11751154
Report Number2648035-2021-07659
Device Sequence Number3092008
Product Code KYB
UDI-Device Identifier05050474530034
UDI-Public(01)05050474530034(17)210923(10)CH14579
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K961242
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/23/2021
Device Model NumberPSCST30
Device Catalogue NumberPSCST30
Device Lot NumberCH14579
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/01/2021
Supplement Date Received by Manufacturer05/06/2021
Initial Report FDA Received Date04/30/2021
Supplement Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Z9002, LENS SN: (B)(4)
-
-