MAUDE Adverse Event Report: MEDLINE INDUSTRIES INC. MEDLINE INDUSTRIES, INC.; CATARACT TENAYA SURGERY CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES INC. MEDLINE INDUSTRIES, INC.; CATARACT TENAYA SURGERY CENTER Back to Search Results
Model Number VAL009CATEA
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported, "blue fibers in the custom pack are evident on the instruments within the pack.Fibers have been found in the eye as well." specifically mentioned, "it was inside the eye, on top of the lens and was removed with utrada forceps, (b)(6) 2021 by md." additional emails received by facility representatives, (b)(6), who provided additional information in regards to this reported incident.Reporters confirm incident occurred (b)(6) 2021.Reporter states, "product was not thoroughly inspected prior to use; issue was identified during procedure." "physician noticed lint in patient's eye and on instruments." per initial complaint report, lint removed with utrada forceps by physician that day ((b)(6) 2021).Reporter states, "no serious injury, follow-up care or medical intervention needed." no additional general anesthesia or sedation was needed or provided after the reported incident.Reporter states, "no adverse events reported during post-op phone call." no samples are available for return and evaluation.Photograph sent.No further information has been provided.Due to the reported incident, medical intervention and in an abundance of caution, this medwatch is being filed.If any further relevant information is identified or obtained, a supplemental medwatch will be submitted.
 
Event or Problem Description
It was reported, "blue fibers in the custom pack are evident on the instruments within the pack.Fibers have been found in the eye as well." specifically mentioned, "it was inside the eye, on top of the lens and was removed with utrada forceps, (b)(6) 2021 by md.".
 
Additional Manufacturer Narrative
D9 device available for evaluation -n/a g6 type of report - follow-up.H2 if follow-up what type? additional information.H3 device evaluated by manufacturer - n/a h6 type of investigation- 3331, 4110, 4112, 4119 h5 investigation conclusion - zcd00005/defect confirmed: unknown responsible party h10 investigation report reads as follows: 05/11/2021 15:26:40 cst ((b)(6) )."sample evaluation: a photo was received for our investigation.Photo shows a long fiber / thread found in eye under magnification.Cannot identify the particulate is blue.Not certain when photo was taken during or post-surgery.Finished good production records reviewed: review of production assembly instructions shows 20-gauge needle, eye spears and instrument wipe are left inside of vendors packaging.Syringes and tega-derm are placed with components inside of a poly bag.There is no direct contact with drapes, gowns or liner of low lint btc.No cotton towels are used in pack.Without lot information for pack, we cannot pull or review production records.Finished good pack build reviewed: review of the bom shows a low lint btc is used for this pack.Trending:trending was reviewed and there have been 1 additional reported complaint(s) for this issue in the past 6 months.Investigation summary: the account reported finding blue fibers in the pack.The reported issue occurred in item val009catea (lot unknown).Based on the photo and information in this complaint we were able to determine a true root cause as the original source of the particulate could not be identified.".
 
Event or Problem Description
It was reported, "blue fibers in the custom pack are evident on the instruments within the pack.Fibers have been found in the eye as well." specifically mentioned, "it was inside the eye, on top of the lens and was removed with utrada forceps, 3/17/21 by md.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE INDUSTRIES, INC.
Common Device Name
CATARACT TENAYA SURGERY CENTER
Manufacturer (Section D)
MEDLINE INDUSTRIES INC.
three lakes drive
northfield IL 60093
MDR Report Key11752484
Report Number1423395-2021-00021
Device Sequence Number3223694
Product Code OJK
UDI-Device Identifier10889942872398
UDI-Public10889942872398
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVAL009CATEA
Device Catalogue NumberVAL009CATEA
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer04/12/2021
Initial Report FDA Received Date04/30/2021
Supplement Report FDA Received Date06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-