MEDOS INTERNATIONAL SARL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE Back to Search Results
Model Number 242043
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Use of Device Problem (1670); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).Investigation summary: the complaint device was received at the service center/supplier and evaluated.There was no allegation of malfunction against the device from the customer, defects were identified during service evaluation.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: distal tip, objective and window damaged.Tube was bent and dented.Image was cloudy/blurry.Minor scratches on the unit.The parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts were identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported by sales rep through service request that the arthsco,4/140_30_qik/strkr hub device had a bent/damaged/dented distal tip and was cloudy.There was no procedure nor patient involvement reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHSCO,4/140_30_QIK/STRKR HUB
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key11752712
Report Number1221934-2021-01404
Device Sequence Number19714549
Product Code HRX
UDI-Device Identifier10886705026234
UDI-Public10886705026234
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K971996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number242043
Device Catalogue Number242043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2021
Initial Report FDA Received Date04/30/2021
Patient Sequence Number1
-
-