• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 NAVIGATION; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S7 NAVIGATION; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733856
Device Problems Imprecision (1307); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9735585, serial/lot #: unknown, if information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used for a cyst fenestration procedure.It was reported that the surgeon felt inaccurate when they were in the brain.The surgeon then noticed that the articulating arm was loose.There was a path in the brain from a previous procedure that the surgeon had the passive catheter introducer aligned to, so the surgeon gently pushed on the articulating arm until navigation on the system aligned with where the surgeon was physically touching in the brain.The inaccuracy was reported to be about.5 mm.There was a less than hour surgical delay and no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 NAVIGATION
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11752915
MDR Text Key248216392
Report Number1723170-2021-01155
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2013
Device Model Number9733856
Device Catalogue Number9733856
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age47 YR
Patient Weight86
-
-