| Model Number |
PED-375-18 |
| Medical Device Problem Codes |
Activation Failure (3270); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the pipeline failed to open distally and became stuck in the marksman catheter hub, which was later found to be accordioned.The patient was undergoing treatment for an unruptured, saccular aneurysm located in the right ophthalmic artery segment.The max diameter was 12mm, and the neck diameter was 7mm.The patient's vessel tortuosity was moderate.The landing zone was 3.36mm distal and 3.68mm proximal.It was reported that after the device was implanted the tip did not open well.Repeated opening caused the microcatheter to rub against the pipeline protective cuff, which could not be separated.The pipeline became stuck at the catheter hub during deployment.A continuous flush had been administered.The physician had released the load/slack in the system, but this did not resolve the issue.It was noted that the proximal segment of the catheter had accordioned.There was no damage to the pushwire.The patient did not experience any symptoms or complications.The patient was noted to be alive with injury as angiographic results post procedure showed a cerebral vascular aneurysm.The devices were prepared according to the instructions for use (ifu).Ancillary devices include a 6f navien guide catheter, marksman microcatheter, and stryker.014" guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: ime patient code updated to e2403.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the pipeline had not been placed in a bend during deployment.The patient was not injured intra-operation, and was currently recovering well.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Product analysis findings: a pipeline flex embolization device was returned for analysis within a shipping box; within a sealed biohazard pouch and within an opened pipeline flex embolization device inner pouch.The marksman catheter used during the event was not returned.No bends or kinks were found with the pipeline pushwire.The hypotube and ptfe were found to be intact.The proximal bumper, re-sheathing marker and re-sheathing pad were found to be intact.The dps sleeves were found to be intact.The tip coil was found to be intact.The proximal braid end was found to be opened and frayed.The distal braid end was found to be opened and frayed.No other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿failure/incomplete open distal¿ was unable to be confirmed as each braid end of the pipeline flex braid were found to be fully open.Possible factors for failure to open is the result of vessel tortuosity or re-sheathing the device more than the recommended two times.Furthermore, the review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Based on the device analysis and reported information, the customer¿s report of ¿resistance/stuck during delivery¿ was unable to be confirmed as the pipeline flex returned without the marksman catheter used during the event.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary.Based on the investigation conducted resistance can occur during tracking, deployment and re-sheathing of the device in distal and tortuous anatomies.In addition, resistance can occur due to patient vessel tortuosity, failure to maintain a continuous flush, or pipeline is pulled back/torqued during delivery.The vessel anatomy was moderate in tortuosity and a continuous flush was maintained.Therefore, the cause could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|