• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.010
Device Problem Vibration (1674)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.This device was returned for service.However, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.Udi: (b)(4).
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the electric pen drive device experienced vibration.It was further determined that the device received an error at the maintenance station.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC PEN DRIVE 60,000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6107428552
MDR Report Key11755520
MDR Text Key249974968
Report Number8030965-2021-03466
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819167711
UDI-Public07611819167711
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Date Manufacturer Received04/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-