• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA28-110/I20-30
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Infection (1930)
Date of Event 07/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx2 abdominal aortic aneurysm stent system.Approximately three (3) years post initial procedure, the patient presented with a graft infection.The physician elected to treat by converting the patient to open repair, explanting the afx devices and replacing with a dacron tube graft on (b)(6) 2020.The final patient status was not reported.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.There are no other equivalent adverse events/incidents for this lot number existing within the endologix complaint handling system.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records or medical imaging relevant to the adverse event/incident was received by endologix.Due to the absence of relevant medical records and reported medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: b4 awareness date.G2 contact office.G3 report source.G4 awareness date.H6 evaluation result codes; remove 3233.H6 evaluation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key11755583
Report Number2031527-2021-00186
Device Sequence Number16933447
Product Code MIH
UDI-Device Identifier00818009014542
UDI-Public(01)00818009014542(17)190709
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/09/2019
Device Model NumberBEA28-110/I20-30
Device Lot Number1955107-001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/16/2020
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date04/30/2021
Supplement Report FDA Received Date06/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL, SN: (B)(6).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
-
-