Model Number 0086998
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional serial numbers: (b)(4).Additional lot numbers : 221694,504903,504905,id411,r212,id411,505861,r225.
 
Event or Problem Description
A review of the events indicated that fourteen (14) patient samples tested on the echo lumena automated blood bank system produced inaccurate results.A review indicated that ten (10) patient sample tests using the echo lumena automated blood bank system produced three (3) unexpected false negative results for the anti-e antibody; two (2) for the anti-c antibody; two (2) for the anti-jka antibody; one (1) for the anti-fya antibody; two (2) for the anti-k antibody; one (1) for the anti-k antibody and one (1) for the anti-cw antibody.Four (4) instrument malfunctions produced inaccurate abo using the abo forward typing assay based on historical results for the patients; while producing one (1) inaccurate result for anti-a and three (3) for anti-d.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO LUMENA
Common Device Name
AUTOMATED BLOOD BANK SYSTEM, PRODUCT CODE, KSZ
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway dr.
norcross GA 30091 5625
MDR Report Key11755601
Report Number1034569-2021-00014
Device Sequence Number3721393
Product Code KSZ
UDI-Device Identifier10888234002338
UDI-Public10888234002338
Combination Product (Y/N)N
PMA/510(K) Number
BK070016
Number of Events Summarized14
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0086998
Device Catalogue Number0086998
Device Lot NumberE404,R197,ID409,R210,E417,R204
Type of Report(Section G)Periodic
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-