SYNTHES GMBH COMPACT AIR DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.701 |
Device Problems
Leak/Splash (1354); Output below Specifications (3004); Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the failures found during service and repair were confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the compact air drive device had low power.It was further determined that the device failed pretest for check the power with power test bench.It was noted in the service order that the device did not work and was leaking air.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|