| Model Number |
ASHA4830-01 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a knee scope procedure, the wands ambient super multivac displayed error code e8 "wand error" on the quantum ii console.The procedure was completed with a backup device with no further complication.A delay greater than 30 minutes was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.No visual issues were observed.A functional evaluation revealed the controller generated an e7 error when the wand was connected.When used with a bypass box the device generated plasma as intended.A review of customer provided images shows an e8 error message on a controller.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) previous use.(2) disconnecting the wand from the controller before power has been turned on.
|
| |
|
Search Alerts/Recalls
|