It was reported that during a knee procedure, the camera control unit's screen had no display.The procedure was successfully completed without delay using the same device.No patient injury or other complications were reported.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Factors that could have contributed to the reported event include; residue on the cable contacts.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future occurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|