• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. FOUNDATION KNEE; OBSOLETE: INSERT, ULTRA SZ 6 13MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. FOUNDATION KNEE; OBSOLETE: INSERT, ULTRA SZ 6 13MM Back to Search Results
Model Number 380-13-506
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to an infection requiring a washout.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 17.8 years apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices was the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
OBSOLETE: INSERT, ULTRA SZ 6 13MM
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key11758012
Report Number1644408-2021-00372
Device Sequence Number4492276
Product Code JWH
UDI-Device Identifier00888912117227
UDI-Public(01)00888912117227
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K963028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2003
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/28/2007
Device Model Number380-13-506
Device Catalogue Number380-13-206
Device Lot Number813931
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/08/2021
Supplement Date Received by Manufacturer05/28/2021
Initial Report FDA Received Date04/30/2021
Supplement Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
120-01-032, LOT UNKNOWN; 120-01-035 LOT 884241; 201-01-108 LOT 859301; 202-01-106, LOT UNKNOWN; 323-01-106 LOT 877341; 324-01-106, LOT UNKNOWN
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age70 YR
-
-