|
The reported complaint of the autopulse platform (sn (b)(4)) displayed "system error, out of service, revert to manual cpr" error message was confirmed during functional testing.The system error could not be verified in the platform's archive because data could not be downloaded due to a defective processor board, likely attributted to normal wear and tear.The autopulse platform was manufactured in january 2008 and it is 13 years old, well past beyond its expected serviceable life of 5 years.The processor board needs to be replaced to remedy this issue.Upon visual inspection, cracked front enclosure was observed likely due to user mishandling such as a drop.The observed physical damage is not related to the reported complaint.The damaged front enclosure needs to be replaced to address this issue.As part of routine service during testing, the platform was examined, and unrelated to the reported complaint, the encoder drive shaft does not rotate smoothly, exhibits binding and resistance due to a sticky clutch plate.The sticky clutch plate needs deburring to address the issue.The cause for the sticky clutch could be due to normal wear and tear.The impact of a sticky clutch was not severe enough to make the platform non-functional.The platform archive data could not be downloaded, likely due to defective processor board and therefore, unable to verify the error message.The autopulse platform failed initial functional testing due to the "system error, out of service, revert to manual cpr" error message displayed upon powering up the device; thus, confirming the reported complaint.To remedy, the processor board needs to be replaced.Waiting on customer's approval for service repair.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse platform with serial number (b)(4).
|
|
During device check at customer site, a zoll benelux representative observed the autopulse platform (serial #(b)(4)) displayed "system error, out of service, revert to manual cpr " upon powering on.No patient involvement.
|