C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH NICORE¿ GUIDEWIRE Back to Search Results
Model Number 786726
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Codes Micturition Urgency (1871); Pain (1994); Hematuria (2558); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the inlay optima stent was expected to stay in a patient for 356 days, the doctor saw patients who returned to hospital after experiencing kidney pain was found with stent encrusted in less than 06 months.Also stated this happened to minimum 10 patients and stent was being removed.
 
Additional Manufacturer Narrative
The reported event is confirmed as design related.No physical sample was returned; however, a photo sample was submitted.The photo sample showed encrustation/calcification present on the stent.The ureteral stent user manual includes stent encrustation as a potential complication associated with retrograde/antegrade positioning of indwelling ureteral stents.The user manual also provides instructions to check for signs of encrustation periodically.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.A potential root cause for this event could be "material selection".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.As the reported event is confirmed as design-related, a label/pack review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the inlay optima stent was expected to stay in a patient for 356 days, the doctor saw patients who returned to hospital after experiencing kidney pain were found with stent encrusted in less than 06 months.Also stated this happened to minimum 10 patients and stent was being removed.Per follow up via ibc on 03may2021, the patients experienced symptoms like pain, urge to urinate, hematuria, urgent retrieval of the stent in the operation room (or).Therefore, placed another stent from another brand.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH NICORE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11760246
Report Number1018233-2021-02533
Device Sequence Number4662435
Product Code FAD
UDI-Device Identifier10801741015202
UDI-Public(01)10801741015202
Combination Product (Y/N)N
Initial Reporter CountryAG
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2023
Device Model Number786726
Device Catalogue Number786626
Device Lot NumberNGDT1521
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer11/16/2021
Initial Report FDA Received Date05/03/2021
Supplement Report FDA Received Date12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-