| Model Number |
786726 |
| Medical Device Problem Code |
Calcified (1077)
|
| Health Effect - Clinical Codes |
Micturition Urgency (1871); Pain (1994); Hematuria (2558); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the inlay optima stent was expected to stay in a patient for 356 days, the doctor saw patients who returned to hospital after experiencing kidney pain was found with stent encrusted in less than 06 months.Also stated this happened to minimum 10 patients and stent was being removed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is confirmed as design related.No physical sample was returned; however, a photo sample was submitted.The photo sample showed encrustation/calcification present on the stent.The ureteral stent user manual includes stent encrustation as a potential complication associated with retrograde/antegrade positioning of indwelling ureteral stents.The user manual also provides instructions to check for signs of encrustation periodically.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.A potential root cause for this event could be "material selection".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.As the reported event is confirmed as design-related, a label/pack review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the inlay optima stent was expected to stay in a patient for 356 days, the doctor saw patients who returned to hospital after experiencing kidney pain were found with stent encrusted in less than 06 months.Also stated this happened to minimum 10 patients and stent was being removed.Per follow up via ibc on 03may2021, the patients experienced symptoms like pain, urge to urinate, hematuria, urgent retrieval of the stent in the operation room (or).Therefore, placed another stent from another brand.No medical intervention was reported.
|
| |
|