• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP POSTLAT DHP-LAT SUPT 3H/LT/75MM-SHORT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP POSTLAT DHP-LAT SUPT 3H/LT/75MM-SHORT; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.117.103
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional device product codes: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the 2.7mm variable angle locking screws would not lock into the fourth most proximal hole of the 2.7mm/3.5mm variable angle lcp posterolateral distal humerus plates with lateral support.Both implants remained in the patient.The 4.0mm cannulated screw short and 4.0mm cannulated screw short thread 64mm were stripped or the cannulated hex screwdriver was stripped.A second set for another screwdriver was opened and the handle broke.Both screws remained in the patient.There was ten (10) minutes of surgical delay.The procedure outcome is unknown.The patient outcome was good.This report involves one (1) 2.7mm/3.5mm va-lcp postlat dhp-lat supt 3h/lt/75mm-short.This is report 5 of 6 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM/3.5MM VA-LCP POSTLAT DHP-LAT SUPT 3H/LT/75MM-SHORT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11760500
Report Number2939274-2021-02190
Device Sequence Number3080910
Product Code HRS
UDI-Device Identifier10886982035882
UDI-Public(01)10886982035882
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120717
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.117.103
Device Catalogue Number02.117.103
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/07/2021
Initial Report FDA Received Date05/03/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.7 VA LCKNG SCR SLF-TPNG T8 SD REC 38.; 2.7/3.5 VA PSTLT DHP-LAT SUPT 3H/L/75-SH.; 2.7/3.5 VA PSTLT DHP-LAT SUPT 3H/L/75-SH.; 4.0MM CANNULATED SCREW-SHORT THREAD 58MM.; 4.0MM CANNULATED SCREW-SHORT THREAD 64MM.; CANNULATED 2.5MM HEXAGONAL SCREWDRIVER.; CANNULATED 2.5MM HEXAGONAL SCREWDRIVER.
Patient Age49 YR
-
-